T1 - Significantly Reduced Blood Pressure Measurement Variability for Both Normotensive and Hypertensive Subjects: Effect of Polynomial Curve Fitting of Oscillometric Pulses
AU - Yang, Fangwei
AU - Chen, Fei
AU - Zhu, Mingping
AU - Chen, Aiqing
AU - Zheng, Dingchang
N1 - This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
PY - 2017/7/12
Y1 - 2017/7/12
N2 - This study aimed to compare within-subject blood pressure (BP) variabilities from different measurement techniques. Cuff pressures from three repeated BP measurements were obtained from 30 normotensive and 30 hypertensive subjects. Automatic BPs were determined from the pulses with normalised peak amplitude larger than a threshold (0.5 for SBP, 0.7 for DBP, and 1.0
for MAP). They were also determined from cuff pressures associated with the above thresholds on a fitted curve polynomial curve of the oscillometric pulse peaks. Finally, the standard deviation (SD) of three repeats and its coefficient of variability (CV) were compared between the two automatic techniques. For the normotensive group, polynomial curve fitting significantly reduced SD
of repeats from 3.6 to 2.5 mmHg for SBP and from 3.7 to 2.1 mmHg for MAP and reduced CV from 3.0% to 2.2% for SBP and from 4.3% to 2.4% for MAP (all 푃 < 0.01). For the hypertensive group, SD of repeats decreased from 6.5 to 5.5 mmHg for SBP and from 6.7 to 4.2 mmHg for MAP, and CV decreased from 4.2% to 3.6% for SBP and from 5.8% to 3.8% for MAP (all 푃 < 0.05). In conclusion, polynomial curve fitting of oscillometric pulses had the ability to reduce automatic BP measurement variability.
AB - This study aimed to compare within-subject blood pressure (BP) variabilities from different measurement techniques. Cuff pressures from three repeated BP measurements were obtained from 30 normotensive and 30 hypertensive subjects. Automatic BPs were determined from the pulses with normalised peak amplitude larger than a threshold (0.5 for SBP, 0.7 for DBP, and 1.0
for MAP). They were also determined from cuff pressures associated with the above thresholds on a fitted curve polynomial curve of the oscillometric pulse peaks. Finally, the standard deviation (SD) of three repeats and its coefficient of variability (CV) were compared between the two automatic techniques. For the normotensive group, polynomial curve fitting significantly reduced SD
of repeats from 3.6 to 2.5 mmHg for SBP and from 3.7 to 2.1 mmHg for MAP and reduced CV from 3.0% to 2.2% for SBP and from 4.3% to 2.4% for MAP (all 푃 < 0.01). For the hypertensive group, SD of repeats decreased from 6.5 to 5.5 mmHg for SBP and from 6.7 to 4.2 mmHg for MAP, and CV decreased from 4.2% to 3.6% for SBP and from 5.8% to 3.8% for MAP (all 푃 < 0.05). In conclusion, polynomial curve fitting of oscillometric pulses had the ability to reduce automatic BP measurement variability.