Quantitative comparison of photoplethysmographic waveform characteristics: Effect of measurement site

Vera Hartmann, Haipeng Liu, Fei Chen, Qian Qiu, Stephen Hughes, Dingchang Zheng

Research output: Contribution to journalArticlepeer-review

111 Citations (Scopus)
57 Downloads (Pure)

Abstract

Introduction: Photoplethysmography (PPG) has been widely used to assess cardiovascular function. However, few studies have comprehensively investigated the effect of measurement site on PPG waveform characteristics. This study aimed to provide a quantitative comparison on this. Methods: Thirty six healthy subjects participated in this study. For each subject, PPG signals were sequentially recorded for 1 min from six different body sites (finger, wrist under (anatomically volar), wrist upper (dorsal), arm, earlobe, and forehead) under both normal and deep breathing patterns. For each body site under a certain breathing pattern, the mean amplitude was firstly derived from recorded PPG waveform which was then normalized to derive several waveform characteristics including the pulse peak time (Tp), dicrotic notch time (Tn), and the reflection index (RI). The effects of breathing pattern and measurement site on the waveform characteristics were finally investigated by the analysis of variance (ANOVA) with post hoc multiple comparisons. Results: Under both breathing patterns, the PPG measurements from the finger achieved the highest percentage of analyzable waveforms for extracting waveform characteristics. There were significant effects of breathing pattern on Tn and RI (larger Tn and smaller RI with deep breathing on average, both p < 0.03). The effects of measurement site on mean amplitude, Tp, Tn, and RI were significant (all p < 0.001). The key results were that, under both breathing patterns, the mean amplitude from finger PPG was significantly larger and its Tp and RI were significantly smaller than those from the other five sites (all p < 0.001, except p = 0.04 for the Tp of “wrist under”), and Tn was only significantly larger than that from the earlobe (both p < 0.05). Conclusion: This study has quantitatively confirmed the effect of PPG measurement site on PPG waveform characteristics (including mean amplitude, Tp, Tn, and RI), providing scientific evidence for a better understanding of the PPG waveform variations between different body sites
Original languageEnglish
Article number198
Number of pages8
JournalFrontiers in Physiology
Volume10
DOIs
Publication statusPublished - 5 Mar 2019
Externally publishedYes

Bibliographical note

This is an openaccess article distributed under the terms of the Creative Commons Attribution
License (CC BY). The use, distribution or reproduction in other forums is permitted,
provided the original author(s) and the copyright owner(s) are credited and that the
original publication in this journal is cited, in accordance with accepted academic
practice. No use, distribution or reproduction is permitted which does not comply
with these terms.

Keywords

  • Breathing pattern
  • Multi-site PPG
  • PPG waveform analysis
  • Photoplethysmography
  • Pulse wave analysis

Fingerprint

Dive into the research topics of 'Quantitative comparison of photoplethysmographic waveform characteristics: Effect of measurement site'. Together they form a unique fingerprint.

Cite this