Evaluating the hip range of motion using goniometer and video tracking methods

Mahshid Yazdi Far, Mohammadreza Yazdifar, Jamaluddin Mahmud, Ibrahim Esat, Mahmoud Chizari

Research output: Contribution to conferencePaper

7 Citations (Scopus)
10 Downloads (Pure)

Abstract

The hip joint's range of motion (ROM) is an important clinical parameter used in diagnosing femoroacetabular impingement (FAI). Along with the use of mechanical devices such as goniometers or inclinometers, motion capture and video tracking has been increasingly deployed to aid the hip joint diagnosis. The goniometer can simply measure the joint angles. However, it has some limitations on allowing the clinician to analyse the ROM at the gate and track the hip joint during walking or maximum squat. Motion capture devices are mainly used to analyse the patient's gait and assess the condition of the joints and bones. This study has used both motion capture and goniometer methods to examine the range of motion of three subject volunteers. The results of both methods have been compared and discussed. The results of the motion capture show flexion of 108±12, adduction (in 90° flexion) of 52±2 and internal rotation (in 90° flexion) of 28±2. For the goniometer results, measurements show flexion of 116±10, adduction (in 90° flexion) of 51±7 and internal rotation (in 90° flexion) of 36±2.
Original languageEnglish
Pages77-82
Number of pages5
DOIs
Publication statusPublished - 2013
EventMalaysian International Tribology Conference 2013 (MITC2013) - Kota Kinabalu, Malaysia
Duration: 18 Nov 201320 Nov 2013

Conference

ConferenceMalaysian International Tribology Conference 2013 (MITC2013)
CountryMalaysia
CityKota Kinabalu
Period18/11/1320/11/13

Fingerprint

Goniometers
Bone

Bibliographical note

Open access under CC BY-NC-ND license

Keywords

  • Hip joint motion
  • femoroacetabular impingement
  • goniometer
  • motion capture

Cite this

Yazdi Far, M., Yazdifar, M., Mahmud, J., Esat, I., & Chizari, M. (2013). Evaluating the hip range of motion using goniometer and video tracking methods. 77-82. Paper presented at Malaysian International Tribology Conference 2013 (MITC2013), Kota Kinabalu, Malaysia. https://doi.org/10.1016/j.proeng.2013.12.150

Evaluating the hip range of motion using goniometer and video tracking methods. / Yazdi Far, Mahshid; Yazdifar, Mohammadreza; Mahmud, Jamaluddin; Esat, Ibrahim; Chizari, Mahmoud.

2013. 77-82 Paper presented at Malaysian International Tribology Conference 2013 (MITC2013), Kota Kinabalu, Malaysia.

Research output: Contribution to conferencePaper

Yazdi Far, M, Yazdifar, M, Mahmud, J, Esat, I & Chizari, M 2013, 'Evaluating the hip range of motion using goniometer and video tracking methods' Paper presented at Malaysian International Tribology Conference 2013 (MITC2013), Kota Kinabalu, Malaysia, 18/11/13 - 20/11/13, pp. 77-82. https://doi.org/10.1016/j.proeng.2013.12.150
Yazdi Far M, Yazdifar M, Mahmud J, Esat I, Chizari M. Evaluating the hip range of motion using goniometer and video tracking methods. 2013. Paper presented at Malaysian International Tribology Conference 2013 (MITC2013), Kota Kinabalu, Malaysia. https://doi.org/10.1016/j.proeng.2013.12.150
Yazdi Far, Mahshid ; Yazdifar, Mohammadreza ; Mahmud, Jamaluddin ; Esat, Ibrahim ; Chizari, Mahmoud. / Evaluating the hip range of motion using goniometer and video tracking methods. Paper presented at Malaysian International Tribology Conference 2013 (MITC2013), Kota Kinabalu, Malaysia.5 p.
@conference{5284c9a8265646d0bed791cd0364ecf7,
title = "Evaluating the hip range of motion using goniometer and video tracking methods",
abstract = "The hip joint's range of motion (ROM) is an important clinical parameter used in diagnosing femoroacetabular impingement (FAI). Along with the use of mechanical devices such as goniometers or inclinometers, motion capture and video tracking has been increasingly deployed to aid the hip joint diagnosis. The goniometer can simply measure the joint angles. However, it has some limitations on allowing the clinician to analyse the ROM at the gate and track the hip joint during walking or maximum squat. Motion capture devices are mainly used to analyse the patient's gait and assess the condition of the joints and bones. This study has used both motion capture and goniometer methods to examine the range of motion of three subject volunteers. The results of both methods have been compared and discussed. The results of the motion capture show flexion of 108±12, adduction (in 90° flexion) of 52±2 and internal rotation (in 90° flexion) of 28±2. For the goniometer results, measurements show flexion of 116±10, adduction (in 90° flexion) of 51±7 and internal rotation (in 90° flexion) of 36±2.",
keywords = "Hip joint motion, femoroacetabular impingement, goniometer, motion capture",
author = "{Yazdi Far}, Mahshid and Mohammadreza Yazdifar and Jamaluddin Mahmud and Ibrahim Esat and Mahmoud Chizari",
note = "Open access under CC BY-NC-ND license; Malaysian International Tribology Conference 2013 (MITC2013) ; Conference date: 18-11-2013 Through 20-11-2013",
year = "2013",
doi = "10.1016/j.proeng.2013.12.150",
language = "English",
pages = "77--82",

}

TY - CONF

T1 - Evaluating the hip range of motion using goniometer and video tracking methods

AU - Yazdi Far, Mahshid

AU - Yazdifar, Mohammadreza

AU - Mahmud, Jamaluddin

AU - Esat, Ibrahim

AU - Chizari, Mahmoud

N1 - Open access under CC BY-NC-ND license

PY - 2013

Y1 - 2013

N2 - The hip joint's range of motion (ROM) is an important clinical parameter used in diagnosing femoroacetabular impingement (FAI). Along with the use of mechanical devices such as goniometers or inclinometers, motion capture and video tracking has been increasingly deployed to aid the hip joint diagnosis. The goniometer can simply measure the joint angles. However, it has some limitations on allowing the clinician to analyse the ROM at the gate and track the hip joint during walking or maximum squat. Motion capture devices are mainly used to analyse the patient's gait and assess the condition of the joints and bones. This study has used both motion capture and goniometer methods to examine the range of motion of three subject volunteers. The results of both methods have been compared and discussed. The results of the motion capture show flexion of 108±12, adduction (in 90° flexion) of 52±2 and internal rotation (in 90° flexion) of 28±2. For the goniometer results, measurements show flexion of 116±10, adduction (in 90° flexion) of 51±7 and internal rotation (in 90° flexion) of 36±2.

AB - The hip joint's range of motion (ROM) is an important clinical parameter used in diagnosing femoroacetabular impingement (FAI). Along with the use of mechanical devices such as goniometers or inclinometers, motion capture and video tracking has been increasingly deployed to aid the hip joint diagnosis. The goniometer can simply measure the joint angles. However, it has some limitations on allowing the clinician to analyse the ROM at the gate and track the hip joint during walking or maximum squat. Motion capture devices are mainly used to analyse the patient's gait and assess the condition of the joints and bones. This study has used both motion capture and goniometer methods to examine the range of motion of three subject volunteers. The results of both methods have been compared and discussed. The results of the motion capture show flexion of 108±12, adduction (in 90° flexion) of 52±2 and internal rotation (in 90° flexion) of 28±2. For the goniometer results, measurements show flexion of 116±10, adduction (in 90° flexion) of 51±7 and internal rotation (in 90° flexion) of 36±2.

KW - Hip joint motion

KW - femoroacetabular impingement

KW - goniometer

KW - motion capture

U2 - 10.1016/j.proeng.2013.12.150

DO - 10.1016/j.proeng.2013.12.150

M3 - Paper

SP - 77

EP - 82

ER -