T1 - Effects of room environment and nursing experience on clinical blood pressure measurement: An observational study
AU - Zhang, Meng
AU - Zhang, Xuemei
AU - Chen, Fei
AU - Dong, Birong
AU - Chen, Aiqing
AU - Zheng, Dingchang
PY - 2017/4
Y1 - 2017/4
N2 - Objective
This study aimed to examine the effects of measurement room environment and nursing experience on the accuracy of manual auscultatory blood pressure (BP) measurement.
Materials and methods
A training database with 32 Korotkoff sounds recordings from the British Hypertension Society was played randomly to 20 observers who were divided into four groups according to the years of their nursing experience (i.e. ≥10 years, 1–9 years, nursing students with frequent training, and those without any medical background; five observers in each group). All the observers were asked to determine manual auscultatory systolic blood pressure (SBP) and diastolic blood pressure (DBP) both in a quiet clinical assessment room and in a noisy nurse station area. This procedure was repeated on another day, yielding a total of four measurements from each observer (i.e. two room environments and two repeated determinations on 2 separate days) for each Korotkoff sound. The measurement error was then calculated against the reference answer, with the effects of room environment and nursing experience of the observer investigated.
Results
Our results showed that there was no statistically significant difference for BPs measured under both quiet and noisy environments (P>0.80 for both SBP and DBP). However, there was a significant effect on the measurement accuracy between the observer groups (P<0.001 for both SBP and DBP). The nursing students performed best with overall SBP and DBP errors of −0.8±2.4 and 0.1±1.8 mmHg, respectively. The SBP measurement error from the nursing students was significantly smaller than that for each of the other three groups (all P<0.001).
Conclusion
Our results indicate that frequent nursing trainings are important for nurses to achieve accurate manual auscultatory BP measurement.
AB - Objective
This study aimed to examine the effects of measurement room environment and nursing experience on the accuracy of manual auscultatory blood pressure (BP) measurement.
Materials and methods
A training database with 32 Korotkoff sounds recordings from the British Hypertension Society was played randomly to 20 observers who were divided into four groups according to the years of their nursing experience (i.e. ≥10 years, 1–9 years, nursing students with frequent training, and those without any medical background; five observers in each group). All the observers were asked to determine manual auscultatory systolic blood pressure (SBP) and diastolic blood pressure (DBP) both in a quiet clinical assessment room and in a noisy nurse station area. This procedure was repeated on another day, yielding a total of four measurements from each observer (i.e. two room environments and two repeated determinations on 2 separate days) for each Korotkoff sound. The measurement error was then calculated against the reference answer, with the effects of room environment and nursing experience of the observer investigated.
Results
Our results showed that there was no statistically significant difference for BPs measured under both quiet and noisy environments (P>0.80 for both SBP and DBP). However, there was a significant effect on the measurement accuracy between the observer groups (P<0.001 for both SBP and DBP). The nursing students performed best with overall SBP and DBP errors of −0.8±2.4 and 0.1±1.8 mmHg, respectively. The SBP measurement error from the nursing students was significantly smaller than that for each of the other three groups (all P<0.001).
Conclusion
Our results indicate that frequent nursing trainings are important for nurses to achieve accurate manual auscultatory BP measurement.