A novel real-time computational framework for detecting catheters and rigid guidewires in cardiac catheterization procedures

YingLiang Ma, Mazen Alhrishy, Srinivas Ananth Narayan, Peter Mountney, Kawal S Rhode

    Research output: Contribution to journalArticlepeer-review

    15 Citations (Scopus)
    122 Downloads (Pure)

    Abstract

    Purpose: Catheters and guidewires are used extensively in cardiac catheterization procedures such as heart arrhythmia treatment (ablation), angioplasty and congenital heart disease treatment. Detecting their positions in fluoroscopic X-ray images is important for several clinical applications, for example, motion compensation, co-registration between 2D and 3D imaging modalities and 3D object reconstruction.
    Methods: For the generalized framework, a multiscale vessel enhancement filter is first used to enhance the visibility of wire-like structures in the X-ray images. After applying adaptive binarization method, the centerlines of wire-like objects were extracted. Finally, the catheters and guidewires were detected as a smooth path which is reconstructed from centerlines of target wire-like objects. In order to classify electrode catheters which are mainly used in electrophysiology procedures, additional steps were proposed. First, a blob detection method, which is embedded in vessel enhancement filter with no additional computational cost, localizes electrode positions on catheters. Then the type of electrode catheters can be recognized by detecting the number of electrodes and also the shape created by a series of electrodes. Furthermore, for detecting guiding catheters or guidewires, a localized machine learning algorithm is added into the framework to distinguish between target wire objects and other wire-like artifacts. The proposed framework were tested on total 10,624 images which are from 102 image sequences acquired from 63 clinical cases.
    Results: Detection errors for the coronary sinus (CS) catheter, lasso catheter ring and lasso catheter body are 0.56 ± 0.28 mm, 0.64 ± 0.36 mm and 0.66 ± 0.32 mm, respectively, as well as success rates of 91.4%, 86.3% and 84.8% were achieved. Detection errors for guidewires and guiding catheters are 0.62 ± 0.48 mm and success rates are 83.5%.
    Conclusion: The proposed computational framework do not require any user interaction or prior models and it can detect multiple catheters or guidewires simultaneously and in real-time. The accuracy of the proposed framework is sub-mm and the methods are robust toward low-dose X-ray fluoroscopic images, which are mainly used during procedures to maintain low radiation dose.
    Original languageEnglish
    Pages (from-to)5066-5079
    Number of pages14
    JournalMedical Physics
    Volume45
    Issue number11
    Early online date16 Sept 2018
    DOIs
    Publication statusPublished - Nov 2018

    Bibliographical note

    This is an open access article under the terms of the Creative Commons Attribution‐NonCommercial License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.

    Copyright © and Moral Rights are retained by the author(s) and/ or other copyright owners. A copy can be downloaded for personal non-commercial research or study, without prior permission or charge. This item cannot be reproduced or quoted extensively from without first obtaining permission in writing from the copyright holder(s). The content must not be changed in any way or sold commercially in any format or medium without the formal permission of the copyright holders.

    Keywords

    • Cardiac catheterization procedures
    • Catheter detection
    • Electrophysiology
    • Guidewire detection

    ASJC Scopus subject areas

    • General Engineering

    Fingerprint

    Dive into the research topics of 'A novel real-time computational framework for detecting catheters and rigid guidewires in cardiac catheterization procedures'. Together they form a unique fingerprint.